Rudolf A. Werner, Susanne Lütje, Bilêl Habacha, Lena Bundschuh, Takahiro Higuchi, Andreas K. Buck, Aleksander Kosmala, Constantin Lapa, Markus Essler, Martin A. Lodge, Kenneth J. Pienta, Mario A. Eisenberger, Mark C. Markowski, Michael A. Gorin, Martin G. Pomper, Steven P. Rowe, Ralph Alexander Bundschuh
- Objectives
We evaluated 18F-DCFPyL test–retest repeatability of uptake in normal organs.
Methods
Twenty-two prostate cancer (PC) patients underwent two 18F-DCFPyL PET scans within 7 days within a prospective clinical trial (NCT03793543). In both PET scans, uptake in normal organs (kidneys, spleen, liver, and salivary and lacrimal glands) was quantified. Repeatability was determined by using within-subject coefficient of variation (wCOV), with lower values indicating improved repeatability.
Results
For SUVmean, repeatability was high for kidneys, spleen, liver, and parotid glands (wCOV, range: 9.0%–14.3%) and lower for lacrimal (23.9%) and submandibular glands (12.4%). For SUVmax, however, the lacrimal (14.4%) and submandibular glands (6.9%) achieved higher repeatability, while for large organs (kidneys, liver, spleen, and parotid glands), repeatability was low (range: 14.1%–45.2%).
Conclusion
We found acceptable repeatability of uptake on 18F-DCFPyL PET for normalObjectives
We evaluated 18F-DCFPyL test–retest repeatability of uptake in normal organs.
Methods
Twenty-two prostate cancer (PC) patients underwent two 18F-DCFPyL PET scans within 7 days within a prospective clinical trial (NCT03793543). In both PET scans, uptake in normal organs (kidneys, spleen, liver, and salivary and lacrimal glands) was quantified. Repeatability was determined by using within-subject coefficient of variation (wCOV), with lower values indicating improved repeatability.
Results
For SUVmean, repeatability was high for kidneys, spleen, liver, and parotid glands (wCOV, range: 9.0%–14.3%) and lower for lacrimal (23.9%) and submandibular glands (12.4%). For SUVmax, however, the lacrimal (14.4%) and submandibular glands (6.9%) achieved higher repeatability, while for large organs (kidneys, liver, spleen, and parotid glands), repeatability was low (range: 14.1%–45.2%).
Conclusion
We found acceptable repeatability of uptake on 18F-DCFPyL PET for normal organs, in particular for SUVmean in the liver or parotid glands. This may have implications for both PSMA-targeted imaging and treatment, as patient selection for radioligand therapy and standardized frameworks for scan interpretation (PROMISE, E-PSMA) rely on uptake in those reference organs.…